Operating Temperature
|
-10°C – 65°C
|
Storage Temperature
|
-25°C – 80°C
|
Available RTD
|
Pt100
(JIS C1640-1989, DIN 43760-1980)
JPt100
(KS C1603-1991, JIS C1604-1981)
|
Range of Temperature Input
|
Pt100:-200.0°C to 600°C
(18.48 to 319.59O)
JPt100:-200.0°C to 600°C
(17.14 to 317.28O)
|
Digital Output
|
Digital converted value (14-bit):
-192–16191 (-8192–8192, 0–16000, -8000–8000)
Digital converted value (16-bit):
0–65000 (-32768–32767, 0–64000, -32000–32000)
(First decimal place value*10)
|
Detection of Broken Wires
|
3 wires for each channel
|
Accuracy
|
±0.1% [Full Scale]
|
Max. Conversion Speed
|
50 ms / 1 channel
|
Number of Temperature Input
|
4 Ch. / 1 module
|
Insulation Type
|
Between input terminal and PLC power: Photo-coupler
Between channels: None
|
Connection Terminal
|
18-point Terminal Block
|
Occupied I/O Inputs
|
16
|
Current Consumption (mA)
|
+5 V
|
50
|
+15 V
|
30
|
-15 V
|
10
|